Recall on Defibtech batteries 06/22/2010
0 Comments
 
Contact:
Technical Support: 
Tel: 1-877-453-4507, 1-203-453-4507
Media Inquiries:
Ray Valek,
Tel: 708-352-8695
ray@defibtech.com

 

FOR IMMEDIATE RELEASE - June 03, 2010 - Guilford, CT – Defibtech, LLC, is initiating a voluntary recall of 5,418 DBP-2800 Battery Packs used in the Lifeline AED® and ReviveR AEDTM (semi-automatic external defibrillators). This recall affects all DBP- 2800 Battery Packs shipped prior to June 4, 2007. In rare instances, when the AED is used with an affected battery pack, the AED may falsely detect an error condition during charging for a shock, then cancel charge and not provide therapy. Defibtech determined the need for this recall after learning of four reports from end users of this malfunction during patient use.

The company has identified recommendations for the end customer to follow until the battery pack has been corrected, which allows the battery pack to remain in service. A copy of these recommendations is being mailed to all affected customers. This customer notification, as well as instructions on determining whether a battery pack is affected, can also be found on the www.defibtech.com/batteryFA1 web page. For additional information regarding this recall, please refer to the above referenced web page, contact your distributor, or contact Defibtech at techsupport@defibtech.com, 1-877-453-4507 or 1- 203-453-4507. 

Defibtech will provide customers with a free battery pack update card to address this issue for all affected battery packs. The correction to the battery pack will be able to be performed at the location where the battery pack is deployed using any DDU-100 series AED and a Defibtech supplied battery pack update card. The battery pack update is expected to be available within the next two weeks. 

The DBP-2800 battery packs affected by this recall have been distributed globally to fire departments, EMS, health clubs, schools, and other organizations. The affected battery packs are used in AEDs which can be identified by the words "Lifeline AED®" and "ReviveR AEDTM" on the front of the device. 

The Food and Drug Administration (FDA) has determined that this action is a Class I recall. Any adverse reactions experienced with the use of this product and/or quality problems should also be reported to the FDA’s MedWatch Program by phone at 1-800- FDA-1088, or on the MedWatch website at www.fda.gov/medwatch2. 
http://www.fda.gov/Safety/Recalls/ucm214859.htm
 


Comments


Your comment will be posted after it is approved.


Leave a Reply

    Author

    View my profile on LinkedIn

    Archives

    September 2011
    August 2011
    July 2011
    April 2011
    March 2011
    February 2011
    January 2011
    December 2010
    August 2010
    June 2010
    May 2010
    March 2010
    January 2010
    October 2009
    September 2009
    August 2009
    July 2009

    Categories

    All
    Bed Bugs
    Cdc
    Emergency
    H1n1
    Health
    HSA/Flex System Changes
    Hurricane Preparedness & Resource Guide
    Recalls

    RSS Feed


Copyright 2009© by East-West Safety Training, Inc. All rights reserved.